Page 26
Jenway Catalogue
PFP7
Part code: 500 701
Part Code Description
500 701 PFP7 industrial flame photometer supplied
with Na, K, Ba, Ca and Li filters, connecting
hoses and clips, compressor plug and drain trap
(230V/50Hz)
Ordering Information
• Designed for industrial analysis
• Supplied with Na, K, Li, Ba and Ca filters
• Low temperature, single channel
• Flame failure safety system
• Operates with propane, butane, natural
gas or LPG
Industrial Flame Photometer
The PFP7 is a low temperature, single channel flame photometer
that is designed for the routine determination of sodium,
potassium, calcium, barium and lithium concentrations. The
flame failure safety system makes these products ideal for use
in industrial and educational environments. The use of fine and
coarse sensitivity controls allows for accurate measurements each
and every time.
Technical Specification
Range
0 to 199.9ppm
Limits of detection Na 0.2ppm
K 0.2ppm
Li 0.25ppm
Ca 15ppm
Ba 30ppm
Reproducibility
<1% coefficient of variation for 20
consecutive samples using 10ppm Na set
to read 50
Linearity
<2% error when 3ppm Na/K and 5ppm
Li are set to read 100
Stability
<2% drift over 5 min when continuously
aspirating 10ppm sample set to 50.0
Specificity
Interference from Na/K and Li equal in
concentration to test element
will be <0.5%
Recorder output
Nominal 1.00V for a reading of 100.0
Services
Electrical
90-125V or 190-250V @50/60Hz*
Air
Moisture and oil free, 6 litre/min @ 14psi
Fuel
Propane, butane, natural gas or LPG
Size (w x d x h), mm 420 x 360 x 300
Weight, kg
8
Key Features
PFP7
Note: This product will only operate with the correct type of
compressor and gas regulator as specified.
* Voltage variants available see page 94
Nebuliser
Part code: 500 019
1...,16,17,18,19,20,21,22,23,24,25 27,28,29,30,31,32,33,34,35,36,...104