Page 6
Jenway Catalogue
6051
Part code: 605 001
Colorimeter
6051
• Wide wavelength range from 400 to 710nm
• Ideal for routine and educational use
• Easy to use and conventional style controls
• Sample flexibility with many size options
• Mains or battery operated
Key Features
Part Code Description
605 001 6051 colorimeter supplied with 8 built in filters,
10x10mm cell holder, mains leads and plastic
cuvettes (230V/50Hz)
Ordering Information
The 6051 is a general purpose colorimeter, equipped with easy-
to-use conventional style controls, making it ideal for routine
and educational use.
Standard 10mm square cuvettes, semi-micro volume cuvettes,
flow-through or pour in/suck out cells can all be used with
the standard sample holder. Excellent sampling flexibility
is achieved through the optional cell holder assemblies,
allowing the use of 10, 20, and 40mm cells, whilst multiple
sample handling is possible with the four cell carriage. For
temperature-dependent measurements a heated cell block is
available.
Technical Specification
Wavelength range
400 to 710nm
Wavelength selection 8 in-built gelatin filters: 430, 470,
490, 520, 540, 580, 600 and 710nm
Measurement ranges
0-100%T
0-1.50Abs
0.1 to 1000 Conc
Resolution
1%T, 0.01Abs, 0.1 to 1 Conc
Bandwidth
Typically 40nm
Blank drift
2%/hr after 15 minute warm-up
Photometric linearity 1%T
Light source
Tungsten filament lamp
Output
Analogue, 10mV per digit
Power
230V, 50Hz*
Size (l x w x d), mm 300 x 355 x 120
Weight, kg
3
* Voltage variants available see page 94